Zancha Low-Temperature
TOC
Benefits
Patented reactor design with constant temperature
guarantees effective oxidation
Increased operational safety through application of a single
UV lamp, enabling effective self-monitoring of the system
Sample preparation: automatically cleaned tangential coarse
filter
New, highly precise NDIR detector: Sapphire optical windows
and Kynar® sample cell eliminate corrosion problems
New operating principle of the NDIR detector eliminates the
need for separate IR sources and IR detectors. Thanks to this
design, the system is reliable and requires low maintenance
Available as dual channel
Allows for measurement of grab samples
COD correlation possible
Field of application
Outflow monitoring
Methanol dosing control
Monitoring of drinking water, cooling water and condensation
Wastewater containing salt
Industrial discharge control
Enhanced NDIR Detector
Using NDIR detection of CO2 for TOC analysis is the preferred
method for
complying with U.S. and International standard protocols.
Zancha TOC’s enhanced NDIR detector eliminates the need
for separate infrared
sources and detectors for the sample and reference cells.
This design
provides a sensitive, stable, noise-free environment for accurate
and reliable
CO2 detection.

Zancha TOC Specifications
- Function: Total Organic Carbon (TOC) or Total Carbon
(TC)
- Operating Principle: Ultraviolet-promoted, sodium persulfate
oxidation. Non-dispersive infrared detection
- Organic Measurement Ranges: Unlimited full-scale selectable
ranges between 10 and 10,000 mg/L carbon
- Data Handling: EIA/TIA-232E and CCITT V.28 Serial (unidirectional
)
- Electrical Output: Programmable, user-selectable 4-20mA
or 0-20mA current loop into 1000 ohms maximum
- Alarms: Two programmable alarm levels driving up to 8
programmable Type C* relays
Programmable Outputs: Up to 8 customer programmable outputs
on Type C*relays (4 relays are standard). Can be programmed
to
output any combination of 18 system parameters (includes
the two alarms)
- Display: 4-line, 20 character/line backlit display
- Power Requirements: 115 ± 10% VAC, 50/60 Hz, 2
Amps maximum, 230W
230 ± 10% VAC, 50/60 Hz, 1 Amp maximum, 230W
- Operating Ambient Temperature: > 0°C to 50°C
- Calibration: 2 point liquid, 2 point gas, manual and
automatic
Performance Characteristics**
- Response Time: Less than 8 minutes to T90 at 100 mg/L
configuration TOC
- Repeatability: ± 1% full scale
- Drift: ± 1% of full scale over 72 hours without
calibration at 20°C
- Temperature Stability: Less than 2% of full scale drift
over ambient range of 10° C to 30°C without liquid
calibration
- Inorganic Carbon Removal: > 95% removal when TIC equals
full scale TOC with standard TIC scrubber
Accuracy (% of full scale): ± 1.5% for TOC concentrations
ranging from 0% to 75% of full scale
± 2.5% for TOC concentrations ranging from 75% to
100% of full scale
Sample Requirements
- Inlet Pressure: Atmospheric to 3PSIG (1.00 - 1.21Barr)
- Drain Pressure: Gravity drain, vented to atmosphere
- Suspended Solids: PA2 or PA3 Sample Preparation/Filter
System required where particle size exceeds 200 µm.
3% maximum concentration of suspended solids by volume.
Oxidation rate is dependent on constituency Flow Rate of
Sample
- Through unit: 35 ml/min. maximum at 60 Hz
Carrier Gas Compressed air containing < 3mg/L CO2. Consumption:
Less than 750 cc/min. at 30 PSIG
- Reagents: 1.5 M Na2S2O8 (sodium persulfate) & 10%
v/v H3PO4 (phosphoric acid) both in deionized water.
- Consumption: 5.2 gallons/month each Physical Characteristics
- Enclosure: NEMA 4, IP 65
- External Dimensions: Height: 30" (76 cm) nominal
Width: 24" (61 cm) nominal
Depth: 22" (56 cm) nominal
Weight: 160 lbs. (72.6 kg) nominal
Shipping Weight: 220 lbs. (99.8 kg) nominal